Sign In | Join Free | My rc363.com
China China World Technology Medical Equipment Service Group logo
China World Technology Medical Equipment Service Group
China World Technology Medical Equipment Service Group Win-win together!
Verified Supplier

5 Years

Home > Patient Monitor Parts >

Patient Bedside Monitor ECG Board Mindray MEC2000 PM7000 PM8000 PM9000

China World Technology Medical Equipment Service Group
Trust Seal
Verified Supplier
Credit Check
Supplier Assessment
Contact Now

Patient Bedside Monitor ECG Board Mindray MEC2000 PM7000 PM8000 PM9000

Brand Name : Mindray

Model Number : MEC2000 PM7000 PM8000 PM9000

MOQ : 1

Payment Terms : T/T, Western Union,PayPal

Delivery Time : 3-5 working days

Product name : Mindray Main Board Mother Board

Type : MEC2000 PM7000 PM8000 PM9000

Application : ICU,hospital device, Patient Monitor

Warranty : 90days

Features : Professional, reliable,Economic

Power supply source : Electric

Mode of transport : Express delivery

AC Input Voltage Rate : 100 V ~ 240 V,

Minimum purchase contract : 1 pcs

Languages : Multi-lingual Support

Advantages : Technical engineers with 12 years of experience,Original

Services : Online technical support

Contact Now

Mindray MEC2000 PM7000 PM8000 PM9000 patient monitor ECG board

►Brief Introduction

This module is designed to measure three parameters including ECG, RESP and TEMP.

Patient Bedside Monitor ECG Board Mindray MEC2000 PM7000 PM8000 PM9000

Specifications

Sensitivity 1 V/mv + 5%(reference 10Hz)
Impedance 50 (ohm)
Signal Delay ECG:25ms
Signal Range ± 8 (mV) p-p

Functions

This module uses ECG cables and TEMP sensors to collect ECG, RESP and TEMP signals,

process them and transmit them to the main board through serial port.

ECG Signal Input Circuit

Input protection and filter circuit: receives ECG signals from ECG cables, removes high-frequency interference and prevents the circuit from being damaged by high voltage generated in defibrillation and ESD.

Right leg drive circuit: picks up 50/60Hz common-mode signals in lead wire and feeds them back to patient body; suppresses the common-mode interference in lead wire for the sake of better detecting ECG signals.

Lead Off detection circuit: detects if any ECG lead falls off and transmits relevant message to CPU.

ECG Signal Process Circuit

Differential Amplification circuit: first-order amplifies ECG signals and suppresses common-mode interference at the same time.

Low-pass filter circuit: removes high-frequency interference outside frequency band of ECG signals.

PACE signals are ECG pacing signals, which greatly affect ECG detecting performance.

Therefore PACE suppression circuit is designed to suppress PACE signals in order to better detect ECG signals.

Master AMP/Filter circuit: amplifies and filters ECG signals again and transmits them furthermore into A/D converter.

Pace Detect

Pick PACE signals out of ECG signals and transmit them to CPU.

Temperature Detect Circuit

Receive signals coming from TEMP sensor, amplify and filter the signals and transmit them furthermore into A/D converter.

Carrier Generate Circuit

RESP measurement is based on Impedance method. Respiration causes the changes of thoracic impedances, which feature is taken advantage to modulate the amplitude of high frequency carrier. The modulated signals are then sent into the measuring circuit. This circuit is designed to generate high frequency carrier

RESP Signal Input Circuit

Couples the RESP signals into the detection circuit.

RESP Signal Process Circuit

Preamplifier circuit: amplifies and filters RESP signals;

Detection circuit: picks out the RESP wave modulated in excitation signals;

Level translation circuit: removes DC components in RESP signals;

Master AMP/Filter circuit: amplifies and filters RESP signals again and transmits them furthermore into A/D converter.

A/D

Convert analog signals output from each parameter circuit into digital signals, and send them into CPU part to receive further processing.

CPU System

Realize logic control over all parameter parts and A/D part.

Process data of each parameter;

Communicate with main board.

Power & Signal Isolate Circuit

Realize isolation from external circuit in order to ensure human safety;

Provide power supplies for circuits;

Realize isolating communication between CPU System and main board.

► About Us


We have supply the Parts, Accessories and medical equipment of Patient Monitor, Defibrillator Machine, ECG Machine, Ventilator, Anesthesia Machine, Ultrasound such as

  • GE,
  • PHiliphhilip,
  • Mindray,
  • Goldway,
  • Nihon Kohden,
  • Med-tronic,
  • Masima,
  • Drager,
  • Edan,
  • Zoll,

etc.

If you want to explore more details about Mindray MEC2000 PM7000 PM8000 PM9000 Patient Bedside Monitor ECG board, please let us know. Looking forward to receiving your early inquiry.


Product Tags:

Patient Bedside Monitor ECG Board

      

PM7000 Monitor ECG Board

      

PM9000 Monitor ECG Board

      
China Patient Bedside Monitor ECG Board Mindray MEC2000 PM7000 PM8000 PM9000 wholesale

Patient Bedside Monitor ECG Board Mindray MEC2000 PM7000 PM8000 PM9000 Images

Inquiry Cart 0
Send your message to this supplier
 
*From:
*To: China World Technology Medical Equipment Service Group
*Subject:
*Message:
Characters Remaining: (0/3000)
 
Inquiry Cart 0